GUI reference
R&S
®
ZNA
925User Manual 1178.6462.02 ─ 29
Table 5-13: Interface types for external device and address formats
Physical
interface
(connector)
Interface
(protocol)
Address Remarks
LAN VXI-11
<IpAddress>
e.g. 127.0.0.0
Full VISA resource string:
TCPIP[board]::<Address>[::INSTR]
SOCKET
<IpAddress>::<PortNo>
e.g. 127.0.0.0::50000
LAN connection with pure TCP/IP protocol;
refer to your VISA user documentation.
GPIB GPIB0 ...
GPIB9
<Address>
e.g. 20
Full VISA resource string:
GPIB[board]::<Address>[::INSTR]
USB USB-VISA
<ManID>::<ProdID>::<SerialNo>
e.g. 0x0AAD::0x0047::100098
2733 (0x0AAD) is the manufacturer ID of
Rohde & Schwarz.
The serial number is device-specific.
LAN or USB Other Interface-specific, e.g. for SOCKET:
TCPIP0::<IpAddress>::<PortNo>::SOCKET
For USB-VISA:
USB0::<ManID>::<ProdID>::<SerialNo>::INSTR
Use complete VISA resource string.
Remote command:
SYSTem:COMMunicate:RDEVice:GDEVice<Gdev>:DEFine
5.19.9.2 External Generic Device Config dialog
The "External Generic Device Config" is a Multi-channel setup dialog. It allows you to
select channel-specific commands (or rather command sequences) to be sent to the
configured generic devices during channel initialization.
Access: System – [Setup] > "Generic Device" > "Generic Device Config..."
●
If the previously measured channel uses the same command to control a generic
device, the command is not sent again.
●
After sending the command to the device, the analyzer waits for the "Settling Time"
requested in the loaded command file.
●
The execution order among the configured devices is undefined.
Background information
Refer to Chapter 4.7.42, "Generic devices", on page 331.
Setup softtool