EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 1286

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1286User Manual 1173.9163.02 ─ 53
The entries consist of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section Error Messages and Troubleshooting.
Example:
SYST:ERR:ALL?
Query all entries in the error queue. 0,"No error" is returned
if the error queue is empty.
Usage: Query only
SYSTem:ERRor:DISPlay[:REMote]
<Boolean>
Disables or enables the display of information popups for remote control errors. These
popups appear at the bottom of the remote screen and the manual screen.
Note
Display of information popups can be globally disabled/enabled using SYSTem:
ERRor:DISPlay:STATe OFF/ON
For SCPI error -113, Undefined header no tooltip is displayed
Parameters:
<Boolean> ON | OFF - enable/disable display of information popups for
remote control errors
*RST: ON
Example:
SYST:ERR:DISP:STATe ON; REM ON
Switch the display of information popups for remote command
errors on.
FREQ:STAR -1
Generate a Remote Error: -222, "Data out of
range; ..." tooltip.
Manual operation: See "Show Remote Error Info Messages" on page 679
SYSTem:ERRor:DISPlay:INFO <Boolean>
SYSTem:ERRor:DISPlay:WARNings <Boolean>
SYSTem:ERRor:DISPlay:ERRor <Boolean>
Selectively disables or enables the display of information popups for the related event
type (Information, Warning, System Error).
Note that the display of information popups can be globally disabled/enabled using
SYSTem:ERRor:DISPlay:STATe OFF/ON.
Parameters:
<Boolean> ON | OFF - enable/disable display of information popups for the
related event type
*RST: ON
SCPI Command Reference

Table of Contents

Related product manuals