EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 1297

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1297User Manual 1173.9163.02 ─ 53
Parameters:
<Type> OFF | DIAGrams | SCReen
OFF touchscreen (e.g. the remote screen) remains active. You
can use the buttons in "Remote" softtool panels. If you switch
back to manual control, the drag and drop functions (e.g. for
markers) are still available.
DIAGrams lock the drag and drop functions in diagrams .
SCReen lock the entire screen including the remote screen
buttons.
*RST: n/a (*RST does not affect the tochscreen lock set-
ting).
Example:
SYSTem:TSLock SCReen
Lock the entire screen.
Manual operation: See "Enabled / Lock Diagrams / Lock Screen" on page 670
SYSTem:USER:DISPlay:TITLe
<String>
Defines a title for the remote display, i.e. for the screen that is shown at the instrument,
if a remote session is established and the standard display is switched off.
Parameters:
<String> Title string.
A \n in the string starts a new line; see example.
*RST: empty string
Example:
SYST:USER:DISP:TITL "DON'T TOUCH\nRemote test
running..."
Define a title for the remote display.
SYSTem:USER:KEY <Key>[, <Label>]
Labels a user-defined key in the remote display. In the query form the command
returns whether or not a user-defined key was tapped or clicked.
Parameters:
<Key> Number of the user key
0 Delete all user keys and restore the default keys ("Go to
Local", "Display Off").
1 to 8 User key numbers
Range: 0 to 8
*RST: 0
<Label> Label for user key no. 1 to 8 (string variable)
SCPI Command Reference

Table of Contents

Related product manuals