Command Reference
R&S
®
ZNB/ZNBT
1315User Manual 1173.9163.02 ─ 53
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect] <Mode>
Selects a search mode for marker no. <Mk>, which can then be initiated using one of
the CALCulate<Ch>:MARKer<Mk>:SEARch...,
CALCulate<Ch>:MARKer<Mk>:MAXimum or
CALCulate<Ch>:MARKer<Mk>:MINimum functions. The marker must be created
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is not needed except for compatibility with ZVR programs. Use
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute to select a search mode and
at the same time initiate the search. The
CALCulate<Ch>:MARKer<Mk>:SEARch...,
CALCulate<Ch>:MARKer<Mk>:MAXimum or
CALCulate<Ch>:MARKer<Mk>:MINimum functions also select the search mode.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number. For a bandfilter search (BFILter) this numeric
suffix is ignored and may be set to any value because the band-
filter search functions always use markers M1 to M4.
Parameters:
<Mode> MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
CALCulate<Chn>:MARKer<Mk>:MAXimum
CALCulate<Chn>:MARKer<Mk>:MINimum
These commands select a search mode for marker no. <Mk> and initiate a maximum
and minimum search, respectively. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This commands are the ZVR-compatible equivalent of CALCulate<Chn>:
MARKer<Mk>:FUNCtion:EXECute MINimum | MAXimum.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number.
Usage: Event
R&S ZVR/ZVABT Compatible Commands