GUI Reference
R&S
®
ZNB/ZNBT
411User Manual 1173.9163.02 ─ 53
analyzer is configured for "Multiple Triggers" (see "FreeRun / External / Manual / Multi-
ple Triggers" on page 409).
Access: CHANNEL – [TRIGGER] > "Trigger Manager..."
Background information
Refer to Chapter 5.1.4.1, "Partial Measurements and Driving Mode", on page 84.
Figure 6-10: Example of a multiple trigger configuration
The table in the "Trigger Manager" dialog contains several editable (white) or non-edit-
able (gray) columns. All settings are analogous to the general trigger settings in the
Trigger Tab. Refer to the following sections:
●
"Sequence" on page 409
●
"FreeRun / External / Manual / Multiple Triggers" on page 409
●
"Signal Type" on page 410
●
"Delay" on page 410
... /Source/ ...
The table defines all settings related to "Multiple Triggers" (see Chapter 6.10.3, "Trig-
ger Tab", on page 407). For all measurement sequences the following trigger sources
are available:
●
"Free Run" means that a trigger signal is not required
●
"External 1" is the external trigger signal fed in at either the EXT TRIG IN connector
on the rear panel or pin 2 of the USER PORT connector.
●
"External 2" is the external trigger signal fed in at pin 25 of the USER PORT con-
nector on the rear panel.
●
"External 1 and 2"/"External 1 or 2" means that the measurement sequence is initi-
ated after the analyzer has received an event from external trigger signal 1 and/or
2. A Signal Type must be specified for both trigger signals.
●
Manual means that the trigger event is generated manually, by selecting the Man-
ual Trigger button.
Sweep Softtool