Remote command:
CONFigure:CHANnel<Ch>[:STATe] OFF
Channel On
Toggles the measurement state of the Active Channel.
Remote command:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Fixture Simulator
The "Fixture Simulator" switch deactivates or activates the configured deembedding,
embedding, balanced ports, and port impedance settings for the selected channel.
When "Fixture Simulator" is deactivated:
all balanced ports are resolved to single ended ports
all port impedances are set to default
all de/embeddings are disabled
At the GUI, the "Balanced Ports" dialog and the de/embedding tabs and dock widgets
are disabled. Related remote commands generate an error.
When "Fixture Simulator" is reactivated, the situation before the deactivation is
restored.
Note: The "Offset" and "One Way Loss" settings are not affected by the "Fixture Simu-
lator" switch. Use the "All Compensation Off"/"All Compensation On" functions of the
"Fixture Compensation" dialog to de/activate these compensations as well (see Chap-
ter 6.13.2.2, "Fixture Compensation Dialog", on page 574).
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:FSIMulator[:STATe]
6.12.1.2 Channel Manager Dialog
The "Channel Manager" dialog allows you to rename channels and to change their
measurement state and sweep mode.
Access: CHANNEL – [CHANNEL CONFIG] > "Channels" > "Channel Manager..."
Background information
Refer to Chapter 5.1.3.3, "Active and Inactive Traces and Channels", on page 83.