781User Manual 1173.9163.02 ─ 53
Return values:
<TestResults>
Usage: Query only
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Mask Test On" on page 632
CALCulate<Chn>:EYE:MASK:FAIL:BEEP <Boolean>
Defines whether the R&S ZNB/ZNBT should make an audible beep on mask failures in
the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Mask Fail Beep" on page 633
CALCulate<Chn>:EYE:MASK:FAIL:CONDition <FailCondition>
Defines whether the fail condition for the eye mask test of the related eye diagram is
specified in absolute or relative terms.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<FailCondition> SAMPles | RATE
SAMPles: The eye mask test will fail if a configurable number of
samples violate the mask (see CALCulate<Chn>:EYE:MASK:
VIOLation:TOLerance).
RATE: The eye mask test will fail if the share of the samples vio-
lating the mask is higher than a configurable percentage (see
CALCulate<Chn>:EYE:MASK:VIOLation:RATE).
*RST: SAMPles
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Test Settings" on page 636