Command Reference
R&S
®
ZNB/ZNBT
800User Manual 1173.9163.02 ─ 53
Parameters:
<Boolean>
Manual operation: See "Automatic Flt Tm for DUT + Lead Ins" on page 590
CALCulate:FMODel:ISD<Ph_pt>:OPERation <Operation>
Defines the execution mode of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Operation> FAST | ACCurate
ACCurate: normal execution mode
FAST: reduces the execution time to ~50%
In many cases the FAST mode can be enabled with only little
loss of accuracy.
Manual operation: See "Operation (fast/acc)" on page 591
CALCulate:FMODel:ISD<Ph_pt>:PASSivity <Boolean>
Defines whether the ISD tool shall enforce passivity and reciprocity for the test cou-
pons and the test fixture.
Corresponds to the passive batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean> ON (0): enforce passivity and reciprocity
OFF (0): do not enforce passivity and reciprocity
Manual operation: See "Enforce Passitivity" on page 591
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer <PortOrder>
Defines how the ISD tool shall interpret the DUT + Fixture data (see CALCulate:
FMODel:ISD<Ph_pt>:DUT:MEASure):
NON: ports 1 to N are on the left and ports N+1 to 2*N are on the right
ODD: ports 1, 3, 5, etc. are on the left and ports 2, 4, 6, etc. are on the right.
ALL: use this to tell the ISD tool that all ports are on the left (i.e. assumed to be
coupled)
Corresponds to the port_order batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<PortOrder> ODD | NON | ALL