Command Reference
R&S
®
ZNB/ZNBT
801User Manual 1173.9163.02 ─ 53
Manual operation: See "Port Sequence" on page 590
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP <String>
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run.
Equivalent to using the ports_to_skip batch mode parameter of the ISD tool with a
list of (positive) port numbers.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String> The port numbers, separated by blanks (e.g. '1 3 4').
Manual operation: See "Ports to Skip (manual)" on page 591
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:LEFT
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:NONE
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:RIGHt
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run.
●
...:LEFT: skip the ports on the left (according to the port order specified using
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer)
●
...:NONE: do not skip any ports
●
...:RIGHt: skip the ports on the right (according to the port order specified using
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer)
Equivalent to setting the ports_to_skip batch mode parameter of the ISD tool to -1,
0 or -2, respectively.
Use CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP to define an arbitrary set of
ports to be skipped.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Ports to Skip" on page 590
CALCulate:FMODel:ISD<Ph_pt>:RUN:RUN
Runs the ISD tool.
Before executing this command, make sure that
●
the test coupon measurement (using CALCulate:FMODel:ISD<Ph_pt>:
COUPon:MEASure:OPEN and CALCulate:FMODel:ISD<Ph_pt>:COUPon:
MEASure:SHORt for 1xOpen+1xShort, CALCulate:FMODel:ISD<Ph_pt>:
COUPon:MEASure otherwise) finished successfully, or the test coupon data were