EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 804

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
804User Manual 1173.9163.02 ─ 53
CALCulate:FMODel:SFD<Ph_pt>:AUTO <Boolean>
Defines whether the SFD tool shall perform automatic impedance adjustments.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
Manual operation: See "Adjust Impedance Mismatch" on page 593
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure
For a fixture modeling with the SFD tool, this command starts the measurement of the
coupon at the active ports (see CALCulate:FMODel:SFD<Ph_pt>:COUPon[:
STATe]).
The coupon type can be selected using CALCulate:FMODel:SFD<Ph_pt>:
COUPon:TYPE.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Measure / Measure Open / Measure Short" on page 585
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure:FILename <String>
For a fixture modeling with the SFD tool, this command loads the coupon properties
from a Touchstone file.
The coupon type can be selected using CALCulate:FMODel:SFD<Ph_pt>:
COUPon:TYPE.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the cur-
rent directory (see MMEMory:CDIRectory)
Manual operation: See "Load File / 1x Open Preset / 1x Short Preset " on page 585
CALCulate:FMODel:SFD<Ph_pt>:COUPon:TYPE <CouponType>
Sets/gets the coupon type to be measured for a fixture modelling with the SFD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<CouponType> SYMMetric2x | OPEN1x | SHORt1x
SCPI Command Reference

Table of Contents

Related product manuals