Command Reference
R&S
®
ZNB/ZNBT
861User Manual 1173.9163.02 ─ 53
CALCulate<Ch>:PARameter:MEASure:SENDed <TraceName>, <Result>
Assigns a measurement result to an existing trace. The query returns the result
assigned to the specified trace (no second parameter; see example).
Similar to CALCulate<Ch>:PARameter:MEASure, but in presence of balanced ports
this command distinguishes between
●
single-ended S-parameters for logical ports (SSS<Li><Lj>)
●
"raw" single-ended S-parameters referring to physical ports (S<Pi><Pj>)
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceName> Trace name
<Result> Measured quantity
Manual operation: See "S-Parameter (selector)" on page 252
CALCulate<Ch>:PARameter:PTIP
<TraceName>, <ProbeTip>
CALCulate<Ch>:PARameter:PTIP? <TraceName>
Gets/sets the probe tip impedance for the calculation of trace <TraceName>.
This is only relevant for Shunt-thru Measurements.
Suffix:
<Ch>
.
Channel number
Parameters:
<ProbeTip> Probe tip impedance
*RST: 0
Default unit: Ohm
Parameters for setting and query:
<TraceName> Trace name
Manual operation: See "Shunt←S21/Probe Tip" on page 272
CALCulate<Ch>:PARameter:SDEFine <TraceName>, <Result>
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace becomes the active trace in the channel but is not displayed.
Note: To display the trace defined via CALCulate<Ch>:PARameter:SDEFine, cre-
ate a diagram (DISPlay[:WINDow<Wnd>][:STATe] ON) and assign the trace to this
diagram (DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED); see example below.
CALCulate<Ch>:PARameter:MEASure changes the measurement result of an
existing trace.
To select an existing trace as the active trace, use CALCulate<Ch>:PARameter:
SELect. You can open the trace manager to obtain an overview of all channels and
traces, including the traces that are not displayed.
SCPI Command Reference