GUI Reference
R&S
®
ZNB/ZNBT
584User Manual 1173.9163.02 ─ 53
Advanced Settings ...
Opens the Advanced Settings dialog of the selected tool. See Chapter 6.13.5.5, "ISD
Advanced Settings", on page 588 and Chapter 6.13.5.6, "SFD Advanced Settings",
on page 592.
Timestamp Filenames
If checked, the names of subsequently generated "Test Coupon" and "DUT + Test Fix-
ture" files are prefixed with the current date and time.
Remote command:
CALCulate:FMODel:REName
Reset to Default
Restores the default settings of the selected fixture modeling tool. This comprises:
●
the states and settings in the Measure Coupon, Measure DUT + Test Fixture and
Save Fixture Model sections (common for all tools)
●
the advanced settings of the selected tool (tool specific)
Remote command:
n.a.
Apply
This button is only active, if the selected deembedding tool has been successfully run
(see Chapter 6.13.5.4, "Save Fixture Model", on page 587).
Applies the calculated deembedding files <...>_left_DUT.sNp and
<...>_right_DUT.sNp) to the ports marked "Active" in the Save Fixture Model sec-
tion.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
Cancel
Closes the dialog without applying any deembedding files.
6.13.5.2 Measure Coupon
The "Measure Coupon" section allows you to measure one or more test coupons for
the related fixture.
Offset Embed Softtool