ZVA/ZVB/ZVT Trace Menu
1145.1084.12 4.49 E-6
and diagram area selected in the drop-down lists. It is possible to create a New channel and/or
d
iagram area for the new trace.
• Delete deletes the selected trace, removing it from the list in the Trace Manager and from the
screen. This button is disabled if the setup contains only one trace: In manual control, each
setup must contain at least one diagram area with one channel and one trace.
Remote
control:
C
ALCulate<Ch>:PARameter:SDEFine <Trace Name>, < Meas
P
arameter>
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED <Trace Name>
CALCulate<Ch>:PARameter:DELete <Trace Name>
Coupling
Selects common channel or scale settings for all traces in the Trace Manager dialog.
The channel and scale coupling is set in two independent panels.
• Decouple All assigns independent channel or scale settings to all traces in the Trace Manager.
If the channel and trace names include numbers, the trace with the lowest number is assigned
to the channel with the lowest number and so forth. Measurement or data traces and their
associated memory traces are assigned to the same channel.
• Couple All assigns all traces to the channel or scale settings selected in the corresponding
drop-down lists. All channel or scale settings except the selected ones are lost. The analyzer
displays a confirmation dialog box before deleting the unused channels.
Remote control:
–
Sort Table
Changes the order of the traces (table rows) in the Trace Manager dialog.
• The drop-down list contains all trace properties that can provide the sorting criterion. The
properties correspond to the column headers in the Trace Manager.