R&S
®
ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.139 E-6
Channel - Stimulus - Stop; see Stimulus and Sweep Types. The time interval between two consecutive
measurements depends on the trigger settings. Any trigger mode is allowed.
The frequency (CW) and internal generator power (Source Power) is fixed and entered into a field which
pops up as soon as CW Mode is activated.
A CW Mode sweep corresponds to the analysis of a signal over the time with a time scale and
resolution that is determined by the trigger events. In a Cartesian diagram the measurement result is
displayed as a trace over a linear time scale (oscillographic representation). The diagram is analogous
to the Time diagram. The following example shows a CW Mode sweep with a DUT that does not
change its characteristics in time:
Sweep time
The time delay between two consecutive trigger pulses must not be smaller than the minimum
measurement time for one measurement point described for Time sweeps.
In CW Mode sweep mode the Start, Center and Stop softkeys are inactive, still you
can use the START CENTER hardkey to select the fixed CW frequency.
Remote control:
[SENSe<Ch>:]SWEep:TPYE POINt
[SENSe<Ch/Tr>:]FUNCtion[:ON] "XTIMe:..."
SOURce<Ch>:FREQuency<Pt>:CW|FIXed
SOURce<Ch>:POWer[:LEVel][:IMMediate][:AMPlitude]
Define Segments
Opens a dialog to define all channel settings for a Segmented Frequency sweep and to import and