ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.161 E-6
Displayed Columns
Opens a dialog to customize the Port Configuration table and display or hide columns.
(Un)checking Logical Port, Source or Receiver displays (removes) the corresponding sections
in (from) the Port Configuration table.
(Un)checking one of the second-level boxes displays (removes) the corresponding column in
(from) the Port Configuration table.
The physical port number is always displayed. The logical port information is displayed only if a
balanced port configuration is defined.
Remote control:
No command, display configuration only.
Balanced and Measured Ports
Opens a dialog to select a balanced port configuration and define reference impedances for balanced
ports and to select the measured ports.
Balanced and Measured Ports
Selecting a balanced port configuration means that one or more pairs of physical test ports are
combined to form logical (balanced) ports.
Balance-unbalance conversion is the simulation of one or more balance-unbalance transformers
(baluns) integrated in the measurement circuit in order to convert the DUT ports from an unbalanced
state into a balanced state and virtually separate the differential and common mode signals. The
analyzer measures the unbalanced state but converts the results and calculates mixed mode
parameters, e.g. mixed mode S-parameters. The balanced ports of the DUT are directly connected to
the analyzer ports and no physical transformer is needed.