Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.176 E-6
RF signal: Fixed port number (1), power sweep range (or fixed power, if the LO signal
is swept), CW frequency.
LO signal: Signal source (analyzer port or external generator), fixed power (or power
sweep range, if the RF signal is at a fixed power), CW frequency.
IF signal: Fixed port number (2), fixed frequency = (RF + LO) or |RF – LO|.
Time or CW mode sweep
For a Time or CW Mode sweep, the signal diagram is labeled as follows:
RF signal: Fixed port number (1), fixed power, CW frequency.
LO signal: Signal source (analyzer port or external generator), fixed power, CW
frequency.
IF signal: Fixed port number (2), fixed frequency = (RF + LO) or |RF – LO|.
Set Power
The Set Power dialog defines the power of the RF and the LO signals.