R&S
®
ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.197 E-6
Start Cal opens a submenu to select a new calibration and start the calibration wizard.
Repeat Previous Cal reopens the wizard to repeat and optimize the previous calibration.
Correction Off activates or deactivates the system error correction in the active channel.
Power Cal opens a submenu controlling the source and receiver power calibration.
Power Correction On (re-)applies the current power calibration in the active channel.
Power Correction Off disables the power calibration in the active channel.
Cal Manager opens a dialog to store system error correction data to a Cal Pool and to assign
correction data to channels.
Recall Last Cal Set loads and activates the setup for which the last calibration was performed.
Cal Kits opens a dialog to manage the calibration kits in use, add new kits and import or export
kits.