R&S
®
ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.199 E-6
One Port P1, One Port P2
Opens a submenu to select a one-port calibration at test ports PORT 1 or PORT 2. The two submenus
are identical:
One-port calibrations can be performed automatically or manually.
• Calibration Unit starts an automatic full one-port calibration. This function is enabled while the
Calibration Unit is connected.
The following menu commands calls up the start dialog of the calibration wizard to start a manual
calibration:
• Normalization (Short) initiates a normalization using a short standard.
• Full initiates a full one-port calibration.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", REFLshort | FOPort , <port_no>
Two Port P1 P2
Opens a submenu to select a two-port calibration at test ports PORT 1 and PORT 2: