Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.200 E-6
Two-port calibrations can be performed automatically or manually.
Calibration Unit starts an automatic tow-port TOSM calibration. This function is enabled while
the Calibration Unit is connected.
The following menu commands calls up the start dialog of the calibration wizard to start a manual
calibration:
Normalization... initiates a normalization using a through standard.
One Path Two Port initiates a full one-port calibration.
TOSM initiates a Through-Open-Short-Match (12-term) calibration
TOM initiates a Through-Open-Match calibration
TRL initiates a Through-Reflect-Line calibration
TNA initiates a Through-Network-Attenuation calibration
Use Start Cal – Other... to select a TOM Enhanced or TSM Enhanced calibration.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", FRTRans | OPTPort | TOSM | TOM | TRL | TNA,
<port_no>,<port_no>
Three Port P1 P2 P3
Opens a submenu to select a three-port TOSM calibration at test ports PORT 1, PORT 2 and PORT 3:
Three-port calibrations can be performed automatically or manually.
Calibration Unit starts an automatic 3-port TOSM calibration. This function is enabled while the
Calibration Unit is connected.
Manual... calls up the start dialog of the calibration wizard to initiate a 3-port TOSM calibration.
Use Start Cal – Other... to select other 3-port calibration types.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)