Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.204 E-6
the calibration standards is mismatched. A standard is assumed to be mismatched if it is
d
escribed by S-parameter data (*.snp file) with a reflection factor > –30 dB.
Remote control:
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect "<Ckit_Name>"
Compile Calibrations
The second dialog of the calibration wizard appears for user-defined (Other...) calibrations only. It is
used to compile a list of calibration types and physical ports in order to perform all calibrations in a
single measurement sequence. A measurement of a particular standard that is required for different
calibration types can be reused.
The selected calibrations and ports appear in the List of Calibrations. A (N) behind the ports number
denotes a node port (see Add Calibration dialog). The list of Standards to be Measured is automatically
generated according to the List of Calibrations.
The dialog provides three buttons to extend or modify the List of Calibrations:
Add Calibration... opens the Add Calibration dialog to add a new calibration to the list.
Modify Calibration... opens the Add Calibration dialog to modify the selected calibration.
Delete Calibration removes the selected calibration from the list.
Two buttons at the bottom of the dialog move back and forward in the calibration wizard:
< Back opens the Select Physical Port Connectors dialog.