ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.205 E-6
> Next opens the Measure Standards dialog.
I
f one of the calibration kits is described by dummy kit parameters or typical values, the analyzer
displays a notice box.
Remote
c
ontrol:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", REFLshort | FOPort | FRTRans | OPTPort | TOSM | TOM
| TRM |
TRL | TNA | ETOM | ETSM,
<port_no>[,<port_no>][,<port_no>][,<port_no>]
Add Calibration
The Add Calibration dialog adds a new calibration to the list in the Compile Calibrations dialog or
modifies an existing calibration.
Type contains a drop-down list to select a calibration type.
Calibrate Ports provides a list of all test ports of the analyzer (no external device ports). Ports
with unsuitable port configuration (see Select ConnectorPhysical Port Number #) are grayed.
Node Port is available for One Path Two Port calibrations where it selects the fully calibrated
port. All ports selected in the Calibrate Ports table can be used as node ports.
Checks on closing the dialog
When the OK buttonis pressed the analyzer checks the calibration kits and the selected calibration (s)
and possibly displays a notice box (confirm with OK) or astatus message. This happens:
If one of the calibration kits is described by dummy kit parameters or typical values.
If one of the selected calibration types is redundant. A calibration is redundant if another
selected calibration provides more complete correction data. Redundant calibrations are
deleted from the list of calibrations.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", REFLshort | FOP
ort | FRTRans | OPTPort | TOSM | TOM
| TRM | TRL | TNA | ETOM | ETSM,
<port_no>[,<port_no>][,<port_no>][,<port_no>]