Display Config R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.250 E-6
Save... opens a Save As... dialog to select a color scheme file and save the current settings.
Color scheme files are non-editable files with the extension *ColorScheme; the default directory
is C:\Rohde&Schwarz\NWA\ColorSchemes.
Recall... opens an Open File... dialog to load and apply a color scheme saved before.
To recall a color scheme file (*.ColorScheme) you can also use the Windows Explorer and simply
double-click the file or drag and drop the file into the NWA application.
Remote
control:
DISPlay:CMAP<Element>:RGB <red>, <green>, <blue>
[,<trace_style>, <trace_width>]
DISPlay:CMAP<Element>:MARKer[:STATe] ON | OFF
DISPlay:CMAP<Element>:TRACe:COLor[:STATe] ON | OFF
MMEMory:LOAD:CMAP
MMEMory:STORe:CMAP
Frequency Info
Shows or hides all frequency stimulus values in the diagrams. This comprises:
The frequency stimulus ranges below the diagram area, if a frequency sweep is active.
The CW frequency in the center below the diagram area, if a power, time or CW mode sweep is
active.
The frequency stimulus values in the marker info field, at the marker position and in the marker
table.
The Frequency Info setting is valid for frequency and segmented frequency sweeps only.
Remote control:
DISPlay:ANNotation:FREQuency[:STATe] ON | OFF