ZVA/ZVB/ZVT System Menu
1145.1084.12 4.253 E-6
Channel Bits
S
ets a channel-dependent four-bit binary value to control four independent output signals at the
USER CONTROL connector (lines 8, 9, 10, 11). The output signals are 3.3 V TTL signals which
can be used to differentiate between up to 16 independent analyzer states. For an application
example refer to the detailed remote control description. Setting the channel bits does not
change the analyzer state.
Resets
Provides several buttons to reset global instrument settings and properties. Global settings (e.g.
the data related to global resources) are not affected by an instrument Preset.
The Use Default Directories button is inactive in the current firmware version.
Option Key Entry
Provides an input field for the option key code supplied with each option. A new software option
for the network analyzer is enabled by entering the option key and pressing Install as described
in the dialog.
Alternatively, it is possible to load the option key from the option key (*.xml) file. Use one of the
following methods:
Copy the *.xml file to the root directory of a floppy disk or memory stick accessible from
your analyzer and press Install.
Copy the *.xml file to the directory C:\Rohde&Schwarz\Nwa of the analyzer's hard disk
and press Install.
You can use several *.xml files in order to enable several options simultaneously. *.xml
files on the storage medium that are no option key files are ignored.
GPIB Settings
Specifies the GPIB Language and an ID String for the analyzer, to be queried via *IDN?.
If the DEFAULT language is activated, the factory ID string Rohde&Schwarz,ZVB<Max.
Freq.-Ports>Port,<Serial_no>,<FW_Version> (e.g. Rohde&Schwarz,ZVB8-
4Port,1145101010100001,1.70.5) is set. The bit order for transferred binary data is
normal (FORMat:BORDer NORMal).
If the PNA language is activated, the an Agilent-compatible ID string is set. The bit
order for transferred binary data is swapped.
The ID string can be changed or reset.
External Power Meters
External Generators
Remote control
(for Messages and
Warnings)
SYSTem:SOUNd:ALARm[:STATe]
SYSTem:SOUNd:STATus[:STATe]
Remote control
(for Keep Measurement
Data...)
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault
Remote control
(for preset and channel
bits):
SYSTem:PRESet:SCOPe SINGle | ALL
CONTrol:AUXiliary:C[:DATA]
OUTPut<Ch>:UPORt[:VALue]