R&S
®
ZVA/ZVB/ZVT TRACe...
1145.1084.12 6.299 E-6
SCPI, Command
Types
Confirmed, no query
Example:
*RST; SWE:POIN 20
Create a trace with 20 sweep points, making the created trace the
active trace of channel 1 (omitted optional mnemonic SENSe1).
CALC:MATH:SDEF 'Trc1 / 2'; CALC:MATH:STAT ON
Define a mathematical trace, dividing the data trace by 2. Activate
the mathematical mode and display the mathematical trace instead
of the data trace.
TRAC:COPY:MATH 'Mem_Pt20',CH1DATA; CALC:MATH:STAT OFF
Copy the current state of the mathematical trace to a memory trace
named "Mem_Pt20". The memory trace is not displayed. Switch the
display back to the data trace.
DISP:WIND:TRAC2:FEED 'MEM_PT20'
Display the created memory trace together with the data trace.
TRACe[:DATA][:RESPonse][:ALL]? CH1DATA | CH2DATA | CH3DATA |
CH4DATA | CH1MEM | CH2MEM | CH3MEM | CH4MEM | MDATA1 | MDATA2 |
MDATA3 | MDATA4 | MDATA5 | MDATA6 | MDATA7 | MDATA8
Returns the response values of the active data trace or memory trace (see trace names).
To read the response values of an arbitrary data or memory trace, use
CALCulate<Ch/Tr>:DATA? To read the response values of a trace acquired in single sweep
mode (INITiate<Ch>:CONTinuous OFF), use CALCulate<Ch/Tr>:DATA:NSWeep?
<Response>
Response data of the selected trace, see list of trace names.
Range [def.
unit]
The data is transferred in the data format defined via FORMat[:DATA].
The unit is the default unit of the measured parameter; see
CALCulate<Ch>:PARameter:SDEFine.
*RST value
SCPI,
Command
Types
Device-specific, query only
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace the
active trace of channel 1 (omitted optional mnemonic SENSe1).
TRAC? CH1DATA
Query the 20 response values of the created trace. In the default
format setting, the data is returned as a comma-separated list of
ASCII values.