EasyManua.ls Logo

Rohde & Schwarz RTA4000 - Page 502

Rohde & Schwarz RTA4000
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTA4000
502User Manual 1335.7898.02 ─ 08
6. For peak searches, the value contains the peak voltage. For
width searches, it contains the pulse width. For edge searches,
the value is not relevant.
Example:
SEARch:RESult:ALL?
Returns all four results of a peak search:
1,-4.7750e-04,0,PEAK,NEGATIVE,-1.530e-02,2,
-4.4630e-04,0,PEAK,NEGATIVE,-1.530e-02,3,
-4.1660e-04,0,PEAK,NEGATIVE,-1.530e-02,4,
-3.8690e-04,0,PEAK,NEGATIVE,-1.530e-02
Example: Chapter 17.2.3.1, "Searching for a pulse of specified width",
on page 429
Usage: Query only
SEARch:RESult<n>?
Returns the result values of the specified search result.
See also: SEARch:RESult:ALL?
Suffix:
<n>
.
*
Number of the search result
Return values:
<Result> Comma-separated value list
Meaning of the values:
Result number, time value, y-position (not relevant), search type,
slope or polarity, optional value: voltage for peak search, pulse
width for width search.
Example:
SEARch:RESult3?
Returns the result values of the third search result.
3,-4.1660e-04,0,PEAK,NEGATIVE,-1.530e-02
Usage: Query only
SEARch:RCOunt?
Returns the number of search results.
Return values:
<ResultCount> *RST: 0
Example: Chapter 17.2.3.1, "Searching for a pulse of specified width",
on page 429
Usage: Query only