Measurements
R&S
®
RTA4000
163User Manual 1335.7898.02 ─ 08
Figure 8-7: Cursor on one source
Figure 8-8: Cursor on two sources. Sources are indicated with the measurement results.
Remote command:
CURSor<m>:USSOURce on page 532
CURSor<m>:SSOURce on page 532
Track Scaling
If enabled, the cursor lines are adjusted when the vertical or horizontal scales are
changed. The cursor lines keep their relative position to the waveform.
If disabled, the cursor lines remain on their position on the display if the scaling is
changed.
Remote command:
CURSor<m>:TRACking:SCALe[:STATe] on page 533
Cursor measurements