Remote control commands
R&S
®
RTA4000
437User Manual 1335.7898.02 ─ 08
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Range> Voltage range value
Range: 8e-3 to 80 (without probe attenuation)
*RST: 40e-3
Default unit: V
CHANnel<m>:POSition <Position>
Sets the vertical position of the waveform in divisions. While the offset sets a voltage,
position is a graphical setting given in divisions.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Position> Graphical position
Range: -5 to 5
*RST: 0
Default unit: div
CHANnel<m>:OFFSet <Offset>
Sets the offset voltage, which is subtracted to correct an offset-affected signal.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Offset> Offset value
Range: Depend on vertical scale and probe attenuation.
Increment: Depends on vertical scale and probe attenuation.
*RST: 0
Default unit: V
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel signal - coupling and termination.
Waveform setup