Fernsteuerbefehle
R&S
®
RTB2000
279Bedienhandbuch 1333.1611.03 ─ 05
CHANnel<m>:RANGe <Range>
Sets the voltage range across the all vertical divisions of the diagram. Use the com-
mand alternatively instead of CHANnel<m>:SCALe.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<Range> Voltage range value
Bereich: 8e-3 bis 80 (without probe attenuation)
*RST: 40e-3
Std.-einheit: V
CHANnel<m>:OFFSet
<Offset>
Sets the offset voltage, which is subtracted to correct an offset-affected signal.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<Offset> Offset value
Bereich: Depend on vertical scale and probe attenuation.
Inkrement: Depends on vertical scale and probe attenuation.
*RST: 0
Std.-einheit: V
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel signal - coupling and termination.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<Coupling> DCLimit | ACLimit | GND
DCLimit
DC coupling passes the input signal unchanged.
ACLimit
Removes the DC offset voltage from the input signal.
GND
Connection to a virtual ground. All channel data is set to o V.
Grundeinstellungen