Fernsteuerbefehle
R&S
®
RTB2000
283Bedienhandbuch 1333.1611.03 ─ 05
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<ThresholdHysteresis>SMALl | MEDium | LARGe
Values correspond to the vertical scale
*RST: SMAL
CHANnel<m>:LABel <Label>
Specifies a name for the selected channel.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<Label> String value
String with max. 8 characters, only ASCII characters can be
used
CHANnel<m>:LABel:STATe
Shows or hides the channel name.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<State> ON | OFF
*RST: OFF
15.4.4 Passive Tastköpfe
PROBe<m>:SETup:ATTenuation:UNIT.............................................................................283
PROBe<m>:SETup:ATTenuation:MANual.........................................................................284
PROBe<m>:SETup:ATTenuation:UNIT <Unit>
Selects the unit that the probe can measure.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.