Fernsteuerbefehle
R&S
®
RTB2000
287Bedienhandbuch 1333.1611.03 ─ 05
Available record length values are:
10 kSa (13107 history segments)
20 kSa (13107 history segments)
50 kSa (3276 history segments)
100 kSa (2621 history segments)
200 kSa (1456 history segments)
500 kSa (319 history segments)
1 MSa (319 history segments)
2 MSa (159 history segments)
5 MSa (40 history segments)
10 MSa (32 history segments)
20 MSa (16 history segments)
Parameter:
<RecordLength> Record length in Samples.
If the entered value differs from the predefined values, the
instrument sets the closest value.
CHANnel<m>:TYPE <DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform
points with lower sample rate.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Parameter:
<DecimationMode> SAMPle | PDETect | HRESolution
SAMPle
Input data is acquired with a sample rate which is aligned to the
time base (horizontal scale) and the record length.
PDETect
Peak Detect: the minimum and the maximum of n samples in a
sample interval are recorded as waveform points.
HRESolution
High resolution: The average of n sample points is recorded as
waveform point.
*RST: SAMPle
ACQuire:TYPE <AcquisitionType>
Sets the type of the aquisition mode.