Fernsteuerbefehle
R&S
®
RTB2000
289Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<HighRes> AUTO | OFF
*RST: OFF
ACQuire:NSINgle:COUNt <NSingleCount>
Sets the number of waveforms acquired with RUNSingle.
Parameter:
<NSingleCount> Number of waveforms.
Bereich: 1 to maximum number that depends on the record
length.
*RST: 1
ACQuire:AVERage:COUNt
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
Parameter:
<AverageCount> Bereich: 2 bis 100,000
*RST: 2
ACQuire:AVERage:RESet
Deletes the waveform and restarts the average calculation.
Verwendung:
Ereignis
ACQuire:AVERage:COMPlete?
Returns the state of averaging.
Rückgabewerte:
<AverageComplete> 0 | 1
0
The number of acquired waveforms is less than the number
required for average calculation. See ACQuire:AVERage:
COUNt auf Seite 289.
1
The instrument acquired a sufficient number of waveforms to
determine the average.
Verwendung: Nur Abfrage
TIMebase:ROLL:AUTomatic <AutomaticRoll>
Enables the automatic roll mode. The instrument switches to roll mode if the timebase
is equal or slower than the roll mode limit defined with TIMebase:ROLL:MTIMe.
Grundeinstellungen