EasyManua.ls Logo

Rohde & Schwarz RTB2000 Series - Page 299

Rohde & Schwarz RTB2000 Series
547 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fernsteuerbefehle
R&S
®
RTB2000
299Bedienhandbuch 1333.1611.03 ─ 05
15.5.5.1 Musterdefinition
TRIGger:A:PATTern:SOURce.......................................................................................... 299
TRIGger:A:PATTern:FUNCtion.........................................................................................299
TRIGger:A:PATTern:CONDition....................................................................................... 299
TRIGger:A:PATTern:SOURce <SourceString>
Sets the state for each channel.
Parameter:
<SourceString> String containing 0, 1, or X for each channel.
1: high, the signal voltage is higher than the trigger level.
0: low, the signal voltage is lower than the trigger level.
X: Don't care. the channel does not affect the trigger.
Without MSO option R&S RTB-B1, the pattern has 4 or 2 bits,
depending on the number of channels: <C1><C2>[<C3><C4>].
With MSO option, the pattern has 20 or 18 bits:
<C1><C2>[<C3><C4>]<D0><D1><D2>...<D15>.
Beispiel: Without MSO option:
TRIG:A:PATT:SOUR "1X10"
C1 and C3 is set to high, C4 to low, and C2 does not matter
(don't care).
Beispiel: With MSO option:
TRIG:A:PATT:SOUR "XXXX111101010011XXXX"
Analog channels C1 to C4 do not matter (don't care). Digital
channels D0 to D11 are set to high or low, D12 to D15 do not
matter.
TRIGger:A:PATTern:FUNCtion <Function>
Sets the logical combination of the channel states .
Parameter:
<Function> AND | OR
AND
The required states of all channels must appear in the input sig-
nal at the same time.
OR
At least one of the channels must have the required state.
*RST: AND
TRIGger:A:PATTern:CONDition <ConditionString>
Defines whether the instrument triggers on fulfillment of the logical condition, or on vio-
lation.
Trigger

Table of Contents