Fernsteuerbefehle
R&S
®
RTB2000
304Bedienhandbuch 1333.1611.03 ─ 05
Suffix:
<m>
.
1..2
Selects the source.
Parameter:
<Source> CH1 | CH2 | CH3 | CH4
Beispiel:
CALC:QMAT:SOUR1 CH2
CALC:QMAT:SOUR2 CH1
CALCulate:QMATh:OPERation <Operation>
Defines the operation that calculates the math waveform
Parameter:
<Operation> ADD | SUB | MUL | DIV
Beispiel:
CALC:QMAT:SOUR1 CH2
CALC:QMAT:SOUR2 CH1
CALC:QMAT:OPER SUB
Subtracts the values of channel 1 from the values of channel 2.
CALCulate:QMATh:POSition <Position>
Sets the vertical position of the math waveform.
Parameter:
<Position> Position value, given in divisions.
CALCulate:QMATh:SCALe <Scale>
Sets the vertical scale of the math waveform.
Parameter:
<Scale> Scale value, given in Volts per division
15.6.3 Referenzmesskurven
In allen REFCurve:...-Befehlen gibt das Suffix <m> die Referenzmesskurve an.
Rufen Sie mit folgenden Befehlen die Messkurvendaten ab:
REFCurve<m>:DATA? auf Seite 382
REFCurve<m>:DATA:HEADer? auf Seite 383
REFCurve<m>:DATA:XINCrement? auf Seite 385
REFCurve<m>:DATA:Xorigin? auf Seite 385
REFCurve<m>:DATA:YINCrement? auf Seite 386
REFCurve<m>:DATA:Yorigin? auf Seite 386
REFCurve<m>:DATA:YRESolution? auf Seite 387
Messkurvenanalyse

Table of Contents