Fernsteuerbefehle
R&S
®
RTB2000
311Bedienhandbuch 1333.1611.03 ─ 05
SEARch:SOURce <SearchSource>
Selects the waveform to be analyzed.
Parameter:
<SearchSource> CH1 | CH2 | CH3 | CH4 | QMA | RE1 | RE2 | RE3 | RE4
Any active channel, math, or reference waveform can be sear-
ched.
For protocol search on CAN and LIN signals, an active serial
bus is the search source.
*RST: CH1
Beispiel: Kapitel 15.2.3.1, "Puls mit angegebener Breite suchen",
auf Seite 271
15.6.4.2
Konfiguration Flankensuche
SEARch:TRIGger:EDGE:SLOPe......................................................................................311
SEARch:TRIGger:EDGE:LEVel........................................................................................311
SEARch:TRIGger:EDGE:LEVel:DELTa............................................................................. 311
SEARch:TRIGger:EDGE:SLOPe <Slope>
Sets the slope to be searched for.
Parameter:
<Slope> POSitive | NEGative | EITHer
*RST: POS
SEARch:TRIGger:EDGE:LEVel <Level>
Sets the voltage level for the edge search.
Parameter:
<Level> *RST: 0.5 V
SEARch:TRIGger:EDGE:LEVel:DELTa <DeltaLevel>
Sets a hysteresis range above and below the search level to avoid unwanted search
results caused by noise oscillation around the level.
Parameter:
<DeltaLevel> Bereich: Lower limit depends on vertical scale and other set-
tings, no upper limit
*RST: 0.2 V