Fernsteuerbefehle
R&S
®
RTB2000
324Bedienhandbuch 1333.1611.03 ─ 05
Each predefined record length corresponds to a maximum number of history seg-
ments, which are stored in the instrument's memory. If option R&S RTB-K15 is instal-
led, you can display the history segments.
Available record length values are:
10 kSa (13107 history segments)
20 kSa (13107 history segments)
50 kSa (3276 history segments)
100 kSa (2621 history segments)
200 kSa (1456 history segments)
500 kSa (319 history segments)
1 MSa (319 history segments)
2 MSa (159 history segments)
5 MSa (40 history segments)
10 MSa (32 history segments)
20 MSa (16 history segments)
Parameter:
<RecordLength> Record length in Samples.
If the entered value differs from the predefined values, the
instrument sets the closest value.
ACQuire:COUNt?
Returns the maximum number of segments that can be captured with the current confi-
guration.
Rückgabewerte:
<NoOfAcquisitions> Number of available segments in the memory
Verwendung: Nur Abfrage
ACQuire:NSINgle:COUNt <NSingleCount>
Sets the number of waveforms acquired with RUNSingle.
Parameter:
<NSingleCount> Number of waveforms.
Bereich: 1 to maximum number that depends on the record
length.
*RST: 1
ACQuire:AVAilable?
Returns the number of segments that are currently saved in the memory. This number
is available for history viewing.