Fernsteuerbefehle
R&S
®
RTB2000
344Bedienhandbuch 1333.1611.03 ─ 05
15.7.2.3 Messbereich
MEASurement<m>:GATE............................................................................................... 344
MEASurement<m>:GATE:MODE.....................................................................................344
MEASurement<m>:GATE:ABSolute:STARt.......................................................................344
MEASurement<m>:GATE:ABSolute:STOP....................................................................... 344
MEASurement<m>:GATE:RELative:STARt.......................................................................345
MEASurement<m>:GATE:RELative:STOP........................................................................345
MEASurement<m>:GATE <State>
Activates or deactivates the measurement gate.
To set the gate, use:
MEASurement<m>:GATE:MODE and
MEASurement<m>:GATE:ABSolute:STARt and MEASurement<m>:GATE:
ABSolute:STOP
or MEASurement<m>:GATE:RELative:STARt and MEASurement<m>:GATE:
RELative:STOP
Suffix:
<m>
.
1..4
Selects the measurement place.
Parameter:
<State> ON | OFF
*RST: OFF
MEASurement<m>:GATE:MODE <GateMode>
Defines whether the gate is defined in absolute or relative values.
Suffix:
<m>
.
1..4
The suffix is irrelevant. One gate is used for all measurement
places.
Parameter:
<GateMode> RELative | ABSolute
*RST: REL
MEASurement<m>:GATE:ABSolute:STARt <Time>
MEASurement<m>:GATE:ABSolute:STOP <Time>
Define the absolute start and end time for the measurement gate, respectively.
Suffix:
<m>
.
1..4
The suffix is irrelevant. One gate is used for all measurement
places.