EasyManua.ls Logo

Rohde & Schwarz RTB2000 Series - Triggerzähler

Rohde & Schwarz RTB2000 Series
547 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fernsteuerbefehle
R&S
®
RTB2000
375Bedienhandbuch 1333.1611.03 ─ 05
Beispiel:
DVM:SOUR CH1
DVM:TYPE MEAN
DVM:RES:STAT?
<-- 4.968E-01,5
The result value of the mean measurement on channel 1 is
496.1 mV. The result status is 5 (decimal) = 0101 (binary). That
means, the result is valid (bit 0 = 1), and the signal is clipped by
the limits of the ADC range (bit 3 = 1).
Verwendung: Nur Abfrage
15.8.6 Triggerzähler
TCOunter:ENABle..........................................................................................................375
TCOunter:SOURce........................................................................................................ 375
TCOunter:RESult[:ACTual]:FREQuency?......................................................................... 375
TCOunter:RESult[:ACTual]:PERiod?................................................................................375
TCOunter:ENABle <Enable>
Enables or disables the trigger counter measurements.
Parameter:
<Enable> ON | OFF
TCOunter:SOURce
Sets the measurement source for the counter.
Parameter:
<CounterSource> CH1 | CH2 | CH3 | CH4 | TRIGger
TRIGger: Trigger source
*RST: TRIG
TCOunter:RESult[:ACTual]:FREQuency?
Returns the frequency of the trigger source.
Rückgabewerte:
<FrequencyValue> Std.-einheit: Hz
Verwendung: Nur Abfrage
TCOunter:RESult[:ACTual]:PERiod?
Returns the period of the trigger source.
Rückgabewerte:
<PeriodValue> Std.-einheit: s
Anwendungen

Table of Contents