Fernsteuerbefehle
R&S
®
RTB2000
381Bedienhandbuch 1333.1611.03 ─ 05
To set the export format, use FORMat[:DATA].
To set the range of samples to be returned, use CHANnel<m>:DATA:POINts.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<EnvelopeData> List of values according to the format settings - the voltages of
the envelope points. The list contains two values for each sam-
ple interval.
Verwendung: Nur Abfrage
CHANnel<m>:DATA:ENVelope:HEADer?
Returns information on the envelope waveform.
Use this command only for envelope waveforms. for all other channel waveforms use
CHANnel<m>:DATA:HEADer?.
Tabelle 15-2: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Number of samples 120000
4 Number of values per sample interval. For envelope
waveforms the value is 2.
2
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameter:
<EnvelopeHeader> Comma-separated value list
Example: -9.477E-008,9.477E-008,200000,2
Verwendung: Nur Abfrage
15.9.1.3 Mathematische Messkurven
Beachten Sie neben den unten beschriebenen Befehlen auch folgende Befehle:
CALCulate:QMATh:DATA:XINCrement? auf Seite 385
CALCulate:QMATh:DATA:XORigin? auf Seite 385
CALCulate:QMATh:DATA:YINCrement? auf Seite 386
CALCulate:QMATh:DATA:YORigin? auf Seite 386