Fernsteuerbefehle
R&S
®
RTB2000
398Bedienhandbuch 1333.1611.03 ─ 05
DISPlay:CLEar[:SCReen]
Deletes all waveforms and measurement results.
Verwendung: Ereignis
DISPlay:PERSistence:TYPE <Type>
Defines how long every new data point remains on the screen.
Parameter:
<Type> OFF | TIME | INFinite
OFF
Deactivates persistence.
TIME
Data points remain on the screen for the duration defined with
DISPlay:PERSistence:TIME.
INF
Data points remain on the screen infinitely until persistence is
set to OFF.
*RST: OFF
DISPlay:PERSistence:TIME <Time>
Persistence time if persistence is active (DISPlay:PERSistence[:STATe] is set to
TIME.
Each new data point in the diagram area remains on the screen for the duration defi-
ned here. .
Parameter:
<Time> Persistence time
Bereich: 50e-3 bis 12.8
Inkrement: 50e-3
*RST: 50e-3
Std.-einheit: s
DISPlay:PERSistence:CLEar
Removes the displayed persistent waveform from the screen.
Verwendung:
Ereignis
DISPlay:PERSistence[:STATe] <State>
Defines whether the waveform persists on the screen or whether the screen is refres-
hed continuously.
Allgemeine Geräteeinstellungen