Fernsteuerbefehle
R&S
®
RTB2000
399Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<State> ON | OFF
ON
The waveform persists for the time defined using DISPlay:
PERSistence:TIME.
OFF
The waveform does not persist on the screen. Only the currently
measured values are displayed.
*RST: OFF
DISPlay:PERSistence:INFinite <InfinitePersistence>
Sets the persistence time to infinite if DISPlay:PERSistence[:STATe] is ON. each
new data point remains on the screen infinitely until this setting is changed or the per-
sistence is cleared.
Parameter:
<InfinitePersistence> ON | OFF
*RST: OFF
DISPlay:DIALog:CLOSe
Closes an open message box.
Verwendung:
Ereignis
DISPlay:DIALog:MESSage <MessageText>
Sends a message text to the instrument and displays it in a message box.
To close the message box, use DISPlay:DIALog:CLOSe.
Einstellparameter:
<MessageText> String
String that contains the message.
Beispiel:
DISP:DIAL:MESS 'My message'
DISP:DIAL:CLOS
Verwendung: Nur Einstellung
DISPlay:GRID:STYLe <Style>
Defines how the grid is displayed.
Allgemeine Geräteeinstellungen