Fernsteuerbefehle
R&S
®
RTB2000
461Bedienhandbuch 1333.1611.03 ─ 05
Suffix:
<b>
.
1 | 2
Parameter:
<BitRate> *RST: 9,6E03
Std.-einheit: Bit/s
15.11.6.2 LIN-Trigger
To configure the protocol trigger, make sure to set first:
TRIGger:A:TYPE to BUS
TRIGger:A:SOURce to SBUS1 | SBUS2
TRIGger:A:LIN:TYPE..................................................................................................... 461
TRIGger:A:LIN:CHKSerror.............................................................................................. 462
TRIGger:A:LIN:IPERror.................................................................................................. 462
TRIGger:A:LIN:SYERror................................................................................................. 462
TRIGger:A:LIN:ICONdition.............................................................................................. 462
TRIGger:A:LIN:IDENtifier................................................................................................ 462
TRIGger:A:LIN:DATA......................................................................................................463
TRIGger:A:LIN:DCONdition.............................................................................................463
TRIGger:A:LIN:DLENgth.................................................................................................463
TRIGger:A:LIN:TYPE <TriggerType>
Specifies the trigger mode for LIN.
Parameter:
<TriggerType> SYNC | WKFRame | ID | IDDT | ERRCondition
SYNC
Start of frame, triggers on the stop bit of the sync field.
WKFRame
Triggers after a wakeup frame.
ID
Sets the trigger to a specific identifier or an identifier range.
Set the identifier with TRIGger:A:LIN:ICONdition and
TRIGger:A:LIN:IDENtifier.
IDDT
Set the identifier (see ID) and the data with TRIGger:A:LIN:
DLENgth, TRIGger:A:LIN:DCONdition and TRIGger:A:
LIN:DATA.
ERRCondition
Identifies various errors in the frame. You can select one or
more error types as trigger condition.
Select the error types with TRIGger:A:LIN:CHKSerror,
TRIGger:A:LIN:IPERror and TRIGger:A:LIN:SYERror.
*RST:
SYNC
Analyse serieller Busse