EasyManua.ls Logo

Rohde & Schwarz RTB2000 Series - Page 471

Rohde & Schwarz RTB2000 Series
547 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fernsteuerbefehle
R&S
®
RTB2000
471Bedienhandbuch 1333.1611.03 ─ 05
SEARch:PROTocol:LIN:ICONdition <IdCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
The command is relevant if SEARch:PROTocol:LIN:CONDition is set to
IDENtifier, IDData or IDERror.
Parameter:
<IdCondition> EQUal | NEQual | GTHan | LTHan
*RST: EQU
SEARch:PROTocol:LIN:IDENtifier
<Identifier>
Defines the identifier pattern.
The command is relevant if SEARch:PROTocol:LIN:CONDition is set to
IDENtifier, IDData or IDERror.
Parameter:
<Identifier> String containing binary pattern. Characters 0, 1, and X are allo-
wed. Enter the 6 bit identifier without parity bits, not the protec-
ted identifier.
SEARch:PROTocol:LIN:DLENgth <DataLength>
Defines the length of the data pattern - the number of bytes in the pattern.
The command is relevant if SEARch:PROTocol:LIN:CONDition is set to IDData.
Parameter:
<DataLength> Bereich: 1 bis 8
Inkrement: 1
*RST: 1
Std.-einheit: Byte
SEARch:PROTocol:LIN:DCONdition <DataCondition>
Sets the comparison condition for data: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
The command is relevant if SEARch:PROTocol:LIN:CONDition is set to IDData.
Parameter:
<DataCondition> EQUal | NEQual | GTHan | LTHan
*RST: EQU