Fernsteuerbefehle
R&S
®
RTB2000
473Bedienhandbuch 1333.1611.03 ─ 05
Rückgabewerte:
<ProbeEnable> 1 | 0
1 = connected, 0 = not connected
Verwendung: Nur Abfrage
LOGic<p>:CURRent:STATe:MAXimum?
DIGital<m>:CURRent:STATe:MAXimum?
LOGic<p>:CURRent:STATe:MINimum?
DIGital<m>:CURRent:STATe:MINimum?
Both commands together return the current status of the indicated logic channel/pod
regardless of the trigger settings, and even without any acquisition.
xxx:CURR:STAT:MIN returns xxx:CURR:STAT:MAX returns Signal
0 0 Low
1 1 High
0 1 Toggle
Suffix:
<m>
.
0..15
Selects the logic channel
<p> 1..2
Selects the logic pod
Rückgabewerte:
<CurrentState> Bereich: 0 | 1
Verwendung: Nur Abfrage
15.12.1.2 Logikanalysator - Konfiguration
LOGic<p>:STATe........................................................................................................... 473
DIGital<m>:DISPlay....................................................................................................... 474
DIGital<m>:TECHnology.................................................................................................474
LOGic<p>:THReshold:UDLevel....................................................................................... 474
LOGic<p>:THReshold.................................................................................................... 474
DIGital<m>:THReshold...................................................................................................475
LOGic<p>:HYSTeresis....................................................................................................475
DIGital<m>:HYSTeresis.................................................................................................. 475
DIGital<m>:SIZE............................................................................................................475
DIGital<m>:POSition...................................................................................................... 476
DIGital<m>:LABel.......................................................................................................... 476
DIGital<m>:LABel:STATe................................................................................................ 476
LOGic<p>:STATe
Switches the logic pod on or off.
Mixed-Signal-Option (Option R&S
RTB-B1)