Fernsteuerbefehle
R&S
®
RTB2000
476Bedienhandbuch 1333.1611.03 ─ 05
DIGital<m>:POSition <Position>
Sets the vertical position of the indicated vertical channel.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameter:
<Position> Vertical position in divisions
Std.-einheit: DIV
DIGital<m>:LABel
<Label>
Defines a label for the indicated logic channel.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameter:
<Label> String value
String parameter
Beispiel:
DIGital4:LABel "Data"
Defines the label "Data" for logic channel D4.
DIGital<m>:LABel:STATe <State>
Displays or hides the label of the indicated logic channel.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameter:
<State> ON | OFF
*RST: OFF
15.12.1.3 Logikkanäle - Messkurvendaten
Beachten Sie für Datenabfragen und -konvertierungen auch folgende Befehle:
FORMat[:DATA] auf Seite 376
DIGital<m>:DATA:XINCrement? auf Seite 385
DIGital<m>:DATA:XORigin? auf Seite 385
DIGital<m>:DATA:YINCrement? auf Seite 386
DIGital<m>:DATA:YORigin? auf Seite 386
DIGital<m>:DATA:YRESolution? auf Seite 387
LOGic<p>:DATA:XINCrement? auf Seite 385
Mixed-Signal-Option (Option R&S
RTB-B1)

Table of Contents