Fernsteuerbefehle
R&S
®
RTB2000
485Bedienhandbuch 1333.1611.03 ─ 05
WGENerator:FREQuency <Frequency>
Defines the frequency value of the selected generator function.
Parameter:
<Frequency> Numeric value depending on the selected generator function
Bereich: 1.000E-01 to 5.000000000E+4 (sine / square)
1.000E-01 to 1.000000000E+4 (pulse / triangle /
ramp)
*RST: 1.00000000E+3
WGENerator:FUNCtion:PULSe:DCYCle
<DutyCycle>
Defines the duty cycle value of the generator function pulse.
Parameter:
<DutyCycle> Numeric value
Bereich: 1.000E+01 bis 9.000E+01
*RST: 2.500E+01
Std.-einheit: %
WGENerator:FUNCtion:PULSe:ETIMe <EdgeTime>
Sets the pulse edge time.
Parameter:
<EdgeTime>
WGENerator:FUNCtion:EXPonential:POLarity <Polarity>
Sets the polarity of the generator function exponential.
Parameter:
<Polarity> POSitive | NEGative
WGENerator:FUNCtion:RAMP:POLarity <Polarity>
Sets the polarity of the generator function ramp.
Parameter:
<Polarity> POSitive | NEGative
*RST: NEG
WGENerator:ARBitrary:SOURce <Source>
Selects the input channel for an arbitrary function to be generated.
Parameter:
<Source> CH1 | CH2 | CH3 | CH4
*RST: CH1