Preparing for Use
R&S
®
RTE
15Getting Started 1326.1026.02 ─ 08
2. Press the [POWER] key on the front panel.
The instrument performs a system check, boots the Windows operating sys-
tem, and then starts the R&S R
TE firmware.
The [POWER] key turns green and the illuminated keys on the front panel
light up. If the previous session was terminated regularly, the oscilloscope
uses the last settings.
To shut down the instrument
► Press the [POWER] key on the front panel.
All current settings are saved, and the software shuts down. The [POWER]
key turns orange. The standby power supplies only the power switch circuits.
Now it is safe to power off the instrument.
The "Exit" function in the "File" menu shuts down only the firmware application. To
shut down the instrument completely, also shut down the operating system in the
"Start" menu, or use the [POWER] key.
3.3.3 EMI Suppression
Electromagnetic Interference (EMI) may affect the measurement results.
T
o suppress generated Electromagnetic Interference:
●
Use suitable shielded cables of high quality. For example use double-shielded
RF and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
3.4 Connecting External Devices
The following interfaces for external devices are provided:
●
USB connectors at the front and rear panel of the instrument
●
Monitor connector DVI-D at the rear panel of the instrument
● Connecting USB Devices................................................................................16
● Connecting an External Monitor......................................................................17
Connecting External Devices