Remote control commands
R&S
®
RTE
1071User Manual 1326.1032.02 ─ 20
:PoWer:DESKew:EXECute;
//Check result
CHANnel2:SKEW:MAN?
CHANnel2:SKEW:TIMe?
POWer:DESKew:TIMe?
*OPC?
Effect of *RST and loading user-defined preset
Note that *RST resets the deskew values.
You can reload the deskew values as follows:
*RST;*OPC?
STOP;*OPC?
:POWer:ENABle
//Select voltage and current sources
:POWer:SOURce:VOLTage1 CHANnel1
:POWer:SOURce:CURRent1 CHANnel2
*OPC?
//Reload deskew values
:POWer:DESKew:CURRent
*OPC?
//Load default saveset after FW restart
MMEM:RCL 'C:\Users\Public\Documents\Rohde-Schwarz\RTx\SaveSets
\UserDefinedPreset_AutoDeskew.dfl'
18.3.7.2 Transient response measurement
Configures and executes a transient response measurement.
Command description in Chapter 18.20.15, "Transient response", on page 1978.
Make sure to configure and deskew the probes before the measurement, see Chap-
ter 18.3.7.1, "Auto deskew", on page 1070.
//Activate power
*RST; *OPC?
:POWer:ENABle
//Expected smps frequency
:POW:TRANsient:FREQ 12500000
:POWer:TRANsient:AUToscale AUTO
:POWer:TRANsient:SIGHigh 0.1
:POWer:TRANsient:SIGLow 0.025
:POWer:TRANsient:HYSTeresis 20
Programming examples