Remote control commands
R&S
®
RTE
1077User Manual 1326.1032.02 ─ 20
ces. The headers of these commands consist of "*" followed by three letters. Many
common commands are related to the Status Reporting System.
Available common commands:
*CAL?.........................................................................................................................1077
*CLS...........................................................................................................................1077
*ESE...........................................................................................................................1078
*ESR?.........................................................................................................................1078
*IDN?..........................................................................................................................1078
*IST?.......................................................................................................................... 1078
*OPC..........................................................................................................................1078
*OPT?.........................................................................................................................1079
*PCB.......................................................................................................................... 1079
*PRE.......................................................................................................................... 1079
*PSC.......................................................................................................................... 1079
*RCL...........................................................................................................................1080
*RST...........................................................................................................................1080
*SAV...........................................................................................................................1080
*SRE.......................................................................................................................... 1080
*STB?.........................................................................................................................1080
*TRG.......................................................................................................................... 1081
*TST?.........................................................................................................................1081
*WAI...........................................................................................................................1081
*CAL?
Starts a self-alignment of the instrument, and then queries a status response. Return
values 0 indicate an error.
Return values:
<State> 0: no error
1: alignment failed
2: not aligned, e.g. init
3: device needs longer warmup time before selfalignment can
start
4: input signal connected during selfalignment
Usage: Query only
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of
the QUEStionable and the OPERation registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
Common commands