Remote control commands
R&S
®
RTE
1086User Manual 1326.1032.02 ─ 20
SYSTem:SHUTdown
Starts the shutdown of the instrument (firmware and operating system).
Usage: Event
SYSTem:PRESet
Resets the instrument to the factory default settings, to the initial state. Factory settings
comprise all instrument settings, including display, intensity and transparency settings.
After loading factory defaults, perform a self-alignment to synchronize the signal data.
Usage: Event
SYSTem:DATE <Year>, <Month>, <Day>
Sets the date of the internal calendar.
Parameters:
<Year> Year, to be entered as a four-digit number (including the century
and millennium information)
<Month> Month, 1 (January) to 12 (December)
<Day> Day, 1 to the maximum number of days in the specified month
*RST: does not affect the date settings
Example:
SYSTem:DATE?
Returned value: 2011,09,13
Usage: SCPI confirmed
SYSTem:TIME? <Hour>, <Minute>, <Second>
Returns the UTC (Universal Time Coordinated) of the internal clock. To define the cur-
rent local time, use the time zone setting of the operating system (SETUP > "Time,
date")
Example:
SYSTem:TIME?
Returned value: 15,09,20. UTC is 15:09:20.
Usage: Query only
SCPI confirmed
SYSTem:DEVice:ID?
Returns the instrument ID - that is the material number and the serial number.
Return values:
<ID> String containing the material number and the serial number
Example:
1316.1000K24-001122-jT
Usage: Query only