Remote control commands
R&S
®
RTE
1163User Manual 1326.1032.02 ─ 20
HDEFinition:RESolution?
Returns the resulting digital resolution in high definition mode.
Return values:
<Resolution> Range: 0 to 16
Increment: 0.1
*RST: 0
Default unit: bit
Usage: Query only
Asynchronous command
FORMat:BORDer
<ByteOrder>
Sets the endianess.
The command is only relevant for raw data export in high definition mode (16 bit word
length).
Parameters:
<ByteOrder> LSBFirst | MSBFirst
LSB first: little endian, least significant byte first
MSB first: big endian, most significant byte first
*RST: LSBFirst
Usage: Asynchronous command
18.8.13 Reference clock
You can select an internal or external reference clock. The external reference is con-
nected to the external reference input on the rear panel.
SENSe[:ROSCillator]:SOURce...................................................................................... 1163
SENSe[:ROSCillator]:EXTernal:FREQuency....................................................................1163
SENSe[:ROSCillator]:SOURce <RefSource>
Enables the use of the external reference signal instead of the internal OCXO refer-
ence.
Parameters:
<RefSource> INTernal | EXTernal
*RST: INTernal
Usage: Asynchronous command
SENSe[:ROSCillator]:EXTernal:FREQuency <ExternalRef>
Sets the frequency of an external reference input signal that is connected to the exter-
nal reference input on the rear panel.
Acquisition and setup